<?xml version="1.0" encoding="UTF-8"?>
<!DOCTYPE article PUBLIC "-//NLM//DTD JATS (Z39.96) Journal Publishing DTD v1.2 20190208//EN" "https://jats.nlm.nih.gov/publishing/1.2/JATS-journalpublishing1.dtd">
<article xmlns:xlink="http://www.w3.org/1999/xlink" xmlns:mml="http://www.w3.org/1998/Math/MathML" article-type="research-article" dtd-version="1.2" xml:lang="en">
<front>
<journal-meta>
<journal-id journal-id-type="publisher-id">IJORS</journal-id>
<journal-title-group><journal-title>International Journal of Ophthalmology and Refractive Surgery</journal-title></journal-title-group>
<publisher><publisher-name>Thomson &amp; Ryberg Publications</publisher-name></publisher>
</journal-meta>
<article-meta>
<article-id pub-id-type="doi">10.62715/ijors.2026.v1.i1.pp21-26</article-id>
<article-id pub-id-type="publisher-id">a056f1e184</article-id>
<article-categories><subj-group subj-group-type="heading"><subject>Methodology Article</subject></subj-group></article-categories>
<title-group><article-title>Standardising Epithelial Thickness Mapping Protocols Before Refractive Surgery Candidate Screening</article-title></title-group>
<contrib-group>
<contrib contrib-type="author" corresp="yes"><name><surname>Watanabe</surname><given-names>Nare</given-names></name><xref ref-type="aff" rid="aff1"><sup>1</sup></xref><xref ref-type="corresp" rid="cor1">*</xref></contrib>
<contrib contrib-type="author"><name><surname>Carrington</surname><given-names>Jinho</given-names></name><xref ref-type="aff" rid="aff2"><sup>2</sup></xref></contrib>
<contrib contrib-type="author"><name><surname>Aziz</surname><given-names>Latoya</given-names></name><xref ref-type="aff" rid="aff3"><sup>3</sup></xref></contrib>
<aff id="aff1"><label>1</label>Department of Medical Sciences, Elbe Institute of Medical Sciences, Oman</aff>
<aff id="aff2"><label>2</label>Department of Medical Sciences, Orinoco Center for Medical Sciences, Czech Republic</aff>
<aff id="aff3"><label>3</label>Department of Medical Sciences, Altai Institute for Advanced Medical Sciences, Bangladesh</aff>
</contrib-group>
<author-notes><corresp id="cor1">* Correspondence: <email>nare.watanabe@elbe.edu</email></corresp></author-notes>
<pub-date pub-type="epub" iso-8601-date="2026-02-22"><day>22</day><month>02</month><year>2026</year></pub-date>
<volume>1</volume>
<issue>1</issue>
<fpage>21</fpage>
<lpage>26</lpage>
<history>
<date date-type="received" iso-8601-date="2025-10-24"><day>24</day><month>10</month><year>2025</year></date>
<date date-type="accepted" iso-8601-date="2025-12-15"><day>15</day><month>12</month><year>2025</year></date>
</history>
<permissions><copyright-statement>© 2026 The Author(s)</copyright-statement><copyright-year>2026</copyright-year><license license-type="open-access" xlink:href="https://creativecommons.org/licenses/by/4.0/"><license-p>This is an open access article distributed under the Creative Commons Attribution 4.0 International License (CC BY 4.0).</license-p></license></permissions>
<self-uri xlink:href="https://trpub.org/article.php?id=a056f1e184"/>
<abstract><p>The corneal epithelium remodels over irregularities of the underlying stroma, thinning over a protrusion and thickening around it. A map of epithelial thickness can therefore reveal early keratoconus that front-surface topography conceals, and can distinguish it from contact lens warpage, which is why such maps are increasingly used when candidates for refractive surgery are screened. The measurement is sensitive to conditions that clinics rarely control: recent lens wear, drops, the tear film, the order of tests, centration, lid position and the device itself. Reported indices and thresholds also vary from one paper and one instrument to the next. This article proposes a protocol in five stages covering preparation of the eye, acquisition, quality control, a minimum reported dataset and interpretation alongside tomography, together with a simple rule for repeating or deferring the examination. Each element is justified from what is known about epithelial behaviour and about the instruments. The protocol is put forward as a proposal. It has not been applied or validated in this work, and the article describes the repeatability and diagnostic accuracy studies that a clinic would need before relying on it or on any cut-off drawn from the literature.</p></abstract>
<kwd-group kwd-group-type="author">
<kwd>Epithelial thickness mapping</kwd>
<kwd>Anterior segment OCT</kwd>
<kwd>Keratoconus screening</kwd>
<kwd>Refractive surgery</kwd>
<kwd>Measurement protocol</kwd>
<kwd>Repeatability</kwd>
</kwd-group>
</article-meta>
</front>
<body>
<sec id="sec1"><title>1. Introduction</title>
<p>Screening before laser refractive surgery is largely a search for the cornea that will not tolerate it. Unrecognised early keratoconus is the strongest known risk factor for ectasia after surgery (Randleman et al., 2008), and the international consensus on ectatic disease treats posterior elevation and thickness distribution, obtained by tomography, as the core of detection (Gomes et al., 2015). Tomography has a blind spot, however. It describes the front surface of the epithelium and the back surface of the cornea, and says nothing directly about the shape of the stromal surface in between.</p>
<p>The epithelium is not a uniform coat. It thins over stromal elevations and thickens over depressions, so that the front surface is smoother than the stroma beneath. In early keratoconus this compensation can hide a cone from topography. Reinstein et al. (2009) showed that the epithelial thickness profile carries the signature of the cone, a localised zone of thinning surrounded by an annulus of thickening, and later work showed that the pattern can be present in fellow eyes that look normal on conventional analysis (Reinstein et al., 2015). Mapping the epithelium has since moved from specialist ultrasound instruments to widely available optical coherence tomography.</p>
<p>What has not followed is agreement on how the map should be acquired and reported. This article proposes a protocol for doing so in the setting of candidate screening. It is a methodological proposal assembled from the literature on epithelial behaviour and instrument performance. No measurements were made for this article and the protocol has not been validated.</p>
</sec>
<sec id="sec2"><title>2. Background</title>
<p>The proposal draws on three bodies of knowledge: the normal profile and its variation, the patterns associated with disease and with lens wear, and the properties of the instruments.</p>
<sec id="sec2-2"><title>2.1. The normal profile</title>
<p>Very high-frequency digital ultrasound first provided maps over a wide diameter. In normal eyes the central epithelium averages a little over 50 micrometres, and it is thinner superiorly than inferiorly, an asymmetry attributed to the action of the upper lid (Reinstein et al., 2008). Optical coherence tomography gave the same qualitative picture and added reference data by sex and age (Kanellopoulos and Asimellis, 2013; Yang et al., 2014). The implication for a protocol is that a normal map is not flat, and that an index of asymmetry must be judged against a normal superior to inferior gradient and not against zero.</p>
</sec>
<sec id="sec2-3"><title>2.2. Patterns that matter in screening</title>
<p>In keratoconus the thinnest epithelium lies over the cone, usually inferotemporal to the centre, with thickening around it; total and stromal thickness maps show the cone more starkly once the epithelium is removed from the calculation (Reinstein et al., 2010). Li et al. (2012) derived pattern-based indices from optical coherence tomography that separated keratoconic from normal eyes, Silverman et al. (2014) built a classifier on epithelial variables, and Temstet et al. (2015) examined the form fruste stage. Contact lens warpage gives the opposite relation between curvature and epithelium: where topography steepens, the epithelium is thick and not thin (Schallhorn et al., 2017). Dry eye has been associated with altered epithelial thickness as well (Kanellopoulos and Asimellis, 2014). A map is therefore read as a pattern in relation to curvature, and a single thickness value has little meaning.</p>
</sec>
<sec id="sec2-4"><title>2.3. Instruments</title>
<p>Optical coherence tomography (Huang et al., 1991) measures the epithelium by segmenting the air to tear interface and the boundary at Bowman&#x27;s layer. The precorneal tear film is included in the optical measurement, whereas immersion ultrasound excludes it. Instruments differ in wavelength, axial resolution, scan pattern, mapped diameter and segmentation software, and their readings are not interchangeable. Repeatability in normal and post-LASIK eyes has been reported as good for a Fourier-domain device (Ma et al., 2013), but precision is poorer toward the periphery and in irregular corneas, which are the eyes of most interest.</p>
</sec>
</sec>
<sec id="sec3"><title>3. Sources of Variability</title>
<p>Table 1 lists the influences that a protocol has to control and the measure proposed for each. They fall into three groups. Some change the epithelium itself, such as recent contact lens wear and ocular surface disease. Some change what the instrument sees without changing the tissue, such as an unstable tear film, residue from viscous drops or a decentred scan. The last group lies in the device and its software, including segmentation errors and differences between instruments.</p>
<table-wrap id="tbl1"><label>Table 1</label><caption><p>Sources of variability in epithelial thickness maps and the control proposed for each</p></caption><table frame="hsides" rules="groups"><thead><tr><th>Source</th><th>Effect on the map</th><th>Proposed control</th></tr></thead><tbody><tr><td>Contact lens wear</td><td>Regional thickening or thinning that mimics or masks disease</td><td>Record lens type and last wear; repeat after a lens-free interval until stable</td></tr><tr><td>Drops and contact tests</td><td>Surface disturbance from anaesthetic, dye, tonometry or gonioscopy</td><td>Map before any drop or contact procedure at the visit</td></tr><tr><td>Tear film</td><td>Optical thickness includes the tear layer; break-up adds noise</td><td>Scan shortly after a natural blink; note surface disease</td></tr><tr><td>Centration and fixation</td><td>Shifts sector values and the apparent site of thinning</td><td>Fix the reference point and reject decentred scans</td></tr><tr><td>Eyelids and lashes</td><td>Missing superior data; pressure distorts the cornea</td><td>Lift the lid against the orbital rim without pressing on the globe</td></tr><tr><td>Segmentation</td><td>Boundary errors produce false focal thin or thick spots</td><td>Inspect the B-scans behind every abnormal sector</td></tr><tr><td>Device and software</td><td>Systematic offsets between instruments and versions</td><td>One device per patient pathway; record model and software version</td></tr></tbody></table></table-wrap>
</sec>
<sec id="sec4"><title>4. The Proposed Protocol</title>
<p>The protocol has five stages, set out in Figure 1 with the element that each stage fixes. It is written to be independent of manufacturer. Where a clinic&#x27;s instrument cannot provide an element, the clinic records that fact instead of substituting a different quantity under the same name.</p>
<fig id="fig1"><label>Figure 1</label><caption><p>Stages of the proposed mapping protocol and the element standardised at each stage</p></caption><graphic xlink:href="IJORS_V1I1_2026_04_Figure1.png" mimetype="image" mime-subtype="png"/></fig>
<sec id="sec4-2"><title>4.1. Stage 1: Preparing the eye</title>
<p>Mapping is placed first in the order of tests, before tonometry, anaesthetic, fluorescein, dilation or any procedure that touches the cornea. Contact lens history is recorded by lens type, wearing pattern and time since last wear. The literature does not support one lens-free interval for all wearers, and the protocol does not invent one. It asks instead for stability: lens wearers are mapped, mapped again after a further lens-free period chosen by the clinic&#x27;s existing policy for refraction and topography, and accepted when the two maps agree within the clinic&#x27;s own repeatability limit. Signs of ocular surface disease are noted, and marked surface disease is treated before the map is used for screening.</p>
</sec>
<sec id="sec4-3"><title>4.2. Stage 2: Acquisition</title>
<p>The same instrument, scan pattern and mapped diameter are used for every candidate, and the widest map the device offers is preferred because the annulus of thickening around a cone may lie outside a small central zone. The reference point for centration, whether pupil centre or corneal vertex, is fixed in the protocol and stated in reports. The patient blinks naturally and the scan is taken within a few seconds. If the upper lid obscures the superior cornea it is lifted against the orbital rim. Three acceptable scans are acquired per eye, with the patient sitting back between them, so that within-session repeatability can be calculated for the individual and not assumed from published averages.</p>
</sec>
<sec id="sec4-4"><title>4.3. Stage 3: Quality control</title>
<p>A scan is accepted only if the manufacturer&#x27;s signal quality index is above the recommended level, the map is complete in all sectors to be reported, centration is within the instrument&#x27;s tolerance and the segmentation lines follow the anatomical boundaries on inspection of the cross-sections. The last check matters most. Automated segmentation can fail over scars, in the periphery and where the tear film is irregular, and a focal abnormality on the map that does not correspond to a visible change in the B-scan is treated as artefact. Rejected scans are counted and the count is recorded.</p>
</sec>
<sec id="sec4-5"><title>4.4. Stage 4: A minimum reported dataset</title>
<p>Published studies report different indices, which makes thresholds hard to compare. The protocol specifies a short common set to be reported from the mean of the accepted scans: central thickness; minimum and maximum thickness with their locations; the difference between minimum and maximum; the superior to inferior difference; the standard deviation across the map; and the mapped diameter. The within-session standard deviation of each index is reported beside it. Indices that depend on proprietary software may be added under their own names. The colour scale and step are fixed so that maps from different visits can be compared by eye.</p>
</sec>
<sec id="sec4-6"><title>4.5. Stage 5: Interpretation with tomography</title>
<p>The map is never read alone. It is placed beside the anterior curvature, posterior elevation and pachymetry maps, and the reader answers one question first: whether the thinnest epithelium coincides with the steepest curvature and the thinnest stroma. Coincident thinning with a surrounding thick zone supports ectasia. Thick epithelium over an area of steepening supports warpage or another non-ectatic cause. A map that is normal in pattern does not clear an eye with abnormal tomography, and an abnormal map in an eye with normal tomography is a reason for further assessment, not a diagnosis.</p>
<p>A repeat or defer rule completes the protocol. If the three scans disagree by more than the clinic&#x27;s repeatability limit, if the pattern is plausibly due to lens wear or surface disease, or if map and tomography conflict, the examination is repeated at a later visit under the same conditions before a surgical decision is taken.</p>
</sec>
</sec>
<sec id="sec5"><title>5. Setting Local Limits</title>
<p>The protocol refers several times to a clinic&#x27;s own repeatability limit, and it does so on purpose. Published cut-offs come from particular instruments, software versions and populations, and carrying them to another setting assumes an equivalence that has not been shown. A clinic adopting the protocol would first establish its own precision, by repeated measurement of a set of normal eyes by the operators who will perform screening, analysed with within-subject standard deviation and limits of agreement in the manner described by Bland and Altman (1986) and, for ophthalmic instruments specifically, by McAlinden et al. (2011). Operator and session effects should be included, since real screening involves both.</p>
</sec>
<sec id="sec6"><title>6. Limitations and What Remains to Be Shown</title>
<p>This article proposes a protocol and reports no test of it. Whether the controls in Table 1 reduce variability in practice, and by how much, is an empirical matter. So is the larger question of whether a standardised map, added to tomography, improves the detection of eyes at risk. Answering it requires a diagnostic accuracy study with a defined reference standard, consecutive enrolment and reporting to the STARD guideline (Bossuyt et al., 2015). The reference standard is itself a difficulty, because early ectasia has no independent gold standard and progression over time is the nearest substitute.</p>
<p>Other limits should be stated. The protocol does not resolve differences between devices and makes no claim that indices transfer from one to another. It does not set a lens-free interval. It addresses screening of untreated eyes; mapping after surgery, where the epithelium remodels in response to the ablation, raises separate questions. Finally, a protocol adds chair time, and a clinic would need to judge whether three scans per eye and inspection of cross-sections are sustainable in its setting.</p>
</sec>
<sec id="sec7"><title>7. Conclusion</title>
<p>Epithelial thickness mapping can show what topography hides, but only if the map reflects the tissue and not the conditions of the examination. The five-stage protocol proposed here fixes the order of tests, the acquisition settings, the quality checks, a minimum set of reported indices and the way the map is read against tomography, and it asks each clinic to derive its own repeatability limits before applying thresholds. It is offered for trial and criticism. Its usefulness will be decided by repeatability and diagnostic accuracy studies that have yet to be carried out.</p>
</sec>
<sec id="sec8"><title>Conflict of interest</title>
<p>To be declared by the author(s) before publication.</p>
</sec>
<sec id="sec9"><title>Funding</title>
<p>To be declared by the author(s) before publication.</p>
</sec>
</body>
<back>
<ref-list><title>References</title>
<ref id="ref1"><label>1</label><mixed-citation publication-type="other"><person-group person-group-type="author"><string-name>Bland, J.M. and Altman, D.G.</string-name></person-group> (<year>1986</year>). <article-title>Statistical methods for assessing agreement between two methods of clinical measurement</article-title>. <source>The Lancet</source>, <volume>327(8476)</volume>, <fpage>307-310</fpage> <ext-link ext-link-type="uri" xlink:href="https://doi.org/10.1016/S0140-6736(86)90837-8">https://doi.org/10.1016/S0140-6736(86)90837-8</ext-link></mixed-citation></ref>
<ref id="ref2"><label>2</label><mixed-citation publication-type="other"><person-group person-group-type="author"><string-name>Bossuyt, P.M., Reitsma, J.B., Bruns, D.E., Gatsonis, C.A., Glasziou, P.P., Irwig, L., Lijmer, J.G., Moher, D., Rennie, D., de Vet, H.C.W., Kressel, H.Y., Rifai, N., Golub, R.M., Altman, D.G., Hooft, L., Korevaar, D.A. and Cohen, J.F.</string-name></person-group> (<year>2015</year>). <article-title>STARD 2015: An updated list of essential items for reporting diagnostic accuracy studies</article-title>. <source>BMJ</source>, <volume>351</volume>, <fpage>h5527</fpage> <ext-link ext-link-type="uri" xlink:href="https://doi.org/10.1136/bmj.h5527">https://doi.org/10.1136/bmj.h5527</ext-link></mixed-citation></ref>
<ref id="ref3"><label>3</label><mixed-citation publication-type="other"><person-group person-group-type="author"><string-name>Gomes, J.A.P., Tan, D., Rapuano, C.J., Belin, M.W., Ambrósio, R. Jr, Guell, J.L., Malecaze, F., Nishida, K. and Sangwan, V.S.</string-name></person-group> (<year>2015</year>). <article-title>Global consensus on keratoconus and ectatic diseases</article-title>. <source>Cornea</source>, <volume>34(4)</volume>, <fpage>359-369</fpage></mixed-citation></ref>
<ref id="ref4"><label>4</label><mixed-citation publication-type="other"><person-group person-group-type="author"><string-name>Huang, D., Swanson, E.A., Lin, C.P., Schuman, J.S., Stinson, W.G., Chang, W., Hee, M.R., Flotte, T., Gregory, K., Puliafito, C.A. and Fujimoto, J.G.</string-name></person-group> (<year>1991</year>). <article-title>Optical coherence tomography</article-title>. <source>Science</source>, <volume>254(5035)</volume>, <fpage>1178-1181</fpage> <ext-link ext-link-type="uri" xlink:href="https://doi.org/10.1126/science.1957169">https://doi.org/10.1126/science.1957169</ext-link></mixed-citation></ref>
<ref id="ref5"><label>5</label><mixed-citation publication-type="other"><person-group person-group-type="author"><string-name>Kanellopoulos, A.J. and Asimellis, G.</string-name></person-group> (<year>2013</year>). <article-title>In vivo three-dimensional corneal epithelium imaging in normal eyes by anterior-segment optical coherence tomography: A clinical reference study</article-title>. <source>Cornea</source>, <volume>32(11)</volume>, <fpage>1493-1498</fpage></mixed-citation></ref>
<ref id="ref6"><label>6</label><mixed-citation publication-type="other"><person-group person-group-type="author"><string-name>Kanellopoulos, A.J. and Asimellis, G.</string-name></person-group> (<year>2014</year>). <article-title>In vivo 3-dimensional corneal epithelial thickness mapping as an indicator of dry eye: Preliminary clinical assessment</article-title>. <source>American Journal of Ophthalmology</source>, <volume>157(1)</volume>, <fpage>63-68</fpage></mixed-citation></ref>
<ref id="ref7"><label>7</label><mixed-citation publication-type="other"><person-group person-group-type="author"><string-name>Li, Y., Tan, O., Brass, R., Weiss, J.L. and Huang, D.</string-name></person-group> (<year>2012</year>). <article-title>Corneal epithelial thickness mapping by Fourier-domain optical coherence tomography in normal and keratoconic eyes</article-title>. <source>Ophthalmology</source>, <volume>119(12)</volume>, <fpage>2425-2433</fpage></mixed-citation></ref>
<ref id="ref8"><label>8</label><mixed-citation publication-type="other"><person-group person-group-type="author"><string-name>Ma, X.J., Wang, L. and Koch, D.D.</string-name></person-group> (<year>2013</year>). <article-title>Repeatability of corneal epithelial thickness measurements using Fourier-domain optical coherence tomography in normal and post-LASIK eyes</article-title>. <source>Cornea</source>, <volume>32(12)</volume>, <fpage>1544-1548</fpage></mixed-citation></ref>
<ref id="ref9"><label>9</label><mixed-citation publication-type="other"><person-group person-group-type="author"><string-name>McAlinden, C., Khadka, J. and Pesudovs, K.</string-name></person-group> (<year>2011</year>). <article-title>Statistical methods for conducting agreement (comparison of clinical tests) and precision (repeatability or reproducibility) studies in optometry and ophthalmology</article-title>. <source>Ophthalmic and Physiological Optics</source>, <volume>31(4)</volume>, <fpage>330-338</fpage></mixed-citation></ref>
<ref id="ref10"><label>10</label><mixed-citation publication-type="other"><person-group person-group-type="author"><string-name>Randleman, J.B., Woodward, M., Lynn, M.J. and Stulting, R.D.</string-name></person-group> (<year>2008</year>). <article-title>Risk assessment for ectasia after corneal refractive surgery</article-title>. <source>Ophthalmology</source>, <volume>115(1)</volume>, <fpage>37-50</fpage></mixed-citation></ref>
<ref id="ref11"><label>11</label><mixed-citation publication-type="other"><person-group person-group-type="author"><string-name>Reinstein, D.Z., Archer, T.J. and Gobbe, M.</string-name></person-group> (<year>2009</year>). <article-title>Corneal epithelial thickness profile in the diagnosis of keratoconus</article-title>. <source>Journal of Refractive Surgery</source>, <volume>25(7)</volume>, <fpage>604-610</fpage></mixed-citation></ref>
<ref id="ref12"><label>12</label><mixed-citation publication-type="other"><person-group person-group-type="author"><string-name>Reinstein, D.Z., Archer, T.J., Gobbe, M., Silverman, R.H. and Coleman, D.J.</string-name></person-group> (<year>2008</year>). <article-title>Epithelial thickness in the normal cornea: Three-dimensional display with Artemis very high-frequency digital ultrasound</article-title>. <source>Journal of Refractive Surgery</source>, <volume>24(6)</volume>, <fpage>571-581</fpage></mixed-citation></ref>
<ref id="ref13"><label>13</label><mixed-citation publication-type="other"><person-group person-group-type="author"><string-name>Reinstein, D.Z., Archer, T.J., Urs, R., Gobbe, M., RoyChoudhury, A. and Silverman, R.H.</string-name></person-group> (<year>2015</year>). <article-title>Detection of keratoconus in clinically and algorithmically topographically normal fellow eyes using epithelial thickness analysis</article-title>. <source>Journal of Refractive Surgery</source>, <volume>31(11)</volume>, <fpage>736-744</fpage></mixed-citation></ref>
<ref id="ref14"><label>14</label><mixed-citation publication-type="other"><person-group person-group-type="author"><string-name>Reinstein, D.Z., Gobbe, M., Archer, T.J., Silverman, R.H. and Coleman, D.J.</string-name></person-group> (<year>2010</year>). <article-title>Epithelial, stromal, and total corneal thickness in keratoconus: Three-dimensional display with Artemis very-high frequency digital ultrasound</article-title>. <source>Journal of Refractive Surgery</source>, <volume>26(4)</volume>, <fpage>259-271</fpage></mixed-citation></ref>
<ref id="ref15"><label>15</label><mixed-citation publication-type="other"><person-group person-group-type="author"><string-name>Schallhorn, J.M., Tang, M., Li, Y., Louie, D.J., Chamberlain, W. and Huang, D.</string-name></person-group> (<year>2017</year>). <article-title>Distinguishing between contact lens warpage and ectasia: Usefulness of optical coherence tomography epithelial thickness mapping</article-title>. <source>Journal of Cataract and Refractive Surgery</source>, <volume>43(1)</volume>, <fpage>60-66</fpage></mixed-citation></ref>
<ref id="ref16"><label>16</label><mixed-citation publication-type="other"><person-group person-group-type="author"><string-name>Silverman, R.H., Urs, R., RoyChoudhury, A., Archer, T.J., Gobbe, M. and Reinstein, D.Z.</string-name></person-group> (<year>2014</year>). <article-title>Epithelial remodeling as basis for machine-based identification of keratoconus</article-title>. <source>Investigative Ophthalmology and Visual Science</source>, <volume>55(3)</volume>, <fpage>1580-1587</fpage></mixed-citation></ref>
<ref id="ref17"><label>17</label><mixed-citation publication-type="other"><person-group person-group-type="author"><string-name>Temstet, C., Sandali, O., Bouheraoua, N., et al.</string-name></person-group> (<year>2015</year>). <article-title>Corneal epithelial thickness mapping using Fourier-domain optical coherence tomography for detection of form fruste keratoconus</article-title>. <source>Journal of Cataract and Refractive Surgery</source>, <volume>41(4)</volume>, <fpage>812-820</fpage></mixed-citation></ref>
<ref id="ref18"><label>18</label><mixed-citation publication-type="other"><person-group person-group-type="author"><string-name>Yang, Y., Hong, J., Deng, S.X. and Xu, J.</string-name></person-group> (<year>2014</year>). <article-title>Age-related changes in human corneal epithelial thickness measured with anterior segment optical coherence tomography</article-title>. <source>Investigative Ophthalmology and Visual Science</source>, <volume>55(8)</volume>, <fpage>5032-5038</fpage></mixed-citation></ref>
</ref-list>
</back>
</article>